Characteristics and Specifications
Vertical system: oscilloscope channels (continued)
Dynamic range
Input impedance
Coupling
BW limit
Channel-to-channel isolation
Standard probes
Probe ID
ESD tolerance
Noise peak-to-peak
DC vertical offset accuracy
1
Single cursor accuracy
Vertical system: logic channels (MSO6000A or MSO-upgraded DSO6000A only)
Number of channels
Threshold groupings
Threshold selections
User-defined threshold range
Maximum input voltage
Input dynamic range
Minimum input voltage swing
Input capacitance
Input resistance
Channel-to-channel skew
1 1 mV/div is a magnification of 2 mV/div setting for 100 MHz models and 2 mV/div is a magnification of 4 mV/div setting for 300 MHz - 1 GHz models. For
vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting and 32 mV for 2 mV/div sensitivity setting.
348
±8 div
MSO/DSO601xA: 1 MΩ ± 1% || 11 pF
MSO/DSO603xA/605xA/610xA: 1 MΩ ± 1% || 14 pF or 50 Ω ± 1.5%, selectable
AC, DC
25 MHz selectable, on the 300 MHz, 500 MHz, and 1 GHz bandwidth models
20 MHz selectable, on the 100 MHz bandwidth models
DC to max bandwidth >40 dB
MSO/DSO601xA: 10:1 10074C shipped standard for each oscilloscope channel
MSO/DSO603xA/605xA/610xA: 10:1 10073C shipped standard for each oscilloscope
channel
MSO/DSO601xA: Auto probe sense
MSO/DSO603xA/605xA/610xA: Auto probe sense and AutoProbe interface
Agilent- and Tektronix-compatible passive probe sense
±2 kV
MSO/DSO601xA: 3% full scale or 2 mV, whichever is greater
MSO/DSO603xA: 3% full scale or 3 mV, whichever is greater
MSO/DSO605xA: 3% full scale or 3.6 mV, whichever is greater
MSO/DSO610xA: 3% full scale or 4 mV, whichever is greater
≤ 200 mV/div: ±0.1 div ±2.0 mV ±0.5% offset value;
>200 mV/div: ±0.1 div ±2.0 mV ±1.5% offset value
±{DC vertical gain accuracy + DC vertical offset accuracy + 0.2% full scale (~1/2 LSB)}
Example: for 50 mV signal, oscilloscope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{2.0% (80 mV) + 0.1 (10 mV) + 2.0 mV + 0.5% (5 mV) + 0.2%(80 mV)} =
± 4.785 mV
16 logic timing channels – labeled D15 - D0
Pod 1: D7 - D0
Pod 2: D15 - D8
TTL, CMOS, ECL and user-definable (selectable by pod)
±8.0 V in 10 mV increments
±40 V peak CAT I; transient overvoltage 800 Vpk
±10 V about threshold
500 mV peak-to-peak
~8 pF
100 kΩ ±2% at probe tip
2 ns typical, 3 ns maximum
6000 Series Oscilloscope User's Guide