6000A Series Oscilloscope Specifications - Agilent Technologies MSO6012A User Manual

6000 series
Hide thumbs Also See for MSO6012A:
Table of Contents

Advertisement

6000A Series Oscilloscope Specifications

Vertical system: oscilloscope channels
Bandwidth (–3dB)
DC vertical gain accuracy
1
Dual cursor accuracy
Vertical system: logic channels (MSO6000A or MSO-upgraded DSO6000A only)
Threshold accuracy
Scope channel triggering
Sensitivity
Logic (D15 - D0) channel triggering (MSO6000A or MSO-upgraded DSO6000A only)
Threshold accuracy
1
1 mV/div is a magnification of 2 mV/div setting for 100 MHz models and 2 mV/div is a magnification of 4 mV/div setting for 300 MHz - 1 GHz models. For
vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting and 32 mV for 2 mV/div sensitivity setting.
6000 Series Oscilloscope User's Guide
All specification are warranted. Specifications are valid after a
30-minute warm-up period and within ±10° C of last "User Cal"
temperature.
Table 19
Warranted specifications
MSO/DSO601xA: DC to 100 MHz
MSO/DSO603xA: DC to 300 MHz
MSO/DSO605xA: DC to 500 MHz
MSO/DSO610xA: DC to 1 GHz
±2.0% full scale
±{DC vertical gain accuracy + 0.4% full scale (~1 LSB)}
Example: for 50 mV signal, oscilloscope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{2.0% (80 mV) + 0.4% (80 mV)} = ±1.92 mV
±(100 mV + 3% of threshold setting)
<10 mV/div: greater of 1 div or 5mV; ≥10 mV/div: 0.6 div
±(100 mV + 3% of threshold setting)