Agilent Technologies MSO6012A User Manual page 105

6000 series
Hide thumbs Also See for MSO6012A:
Table of Contents

Advertisement

2
Front-Panel Controls
To Channel 2
104
For a 4-channel oscilloscope, connect BNC tees to the
equal-length cables as shown below. Then connect a
BNC(f)-to-BNC(f) (barrel connector) to the tee as shown
below.
To Channel 1
Figure 11
User Calibration cable for 4-channel oscilloscope
1 Connect a BNC cable (40 inches maximum) from the TRIG
OUT connector on the rear panel to the BNC barrel
connector.
2 Press the Utility key, then press the Service softkey.
3 Begin the Self Cal by pressing the Start User Cal softkey.
4 When the User Cal is completed, set the rear-panel
CALIBRATION switch to PROTECTED.
User Cal Status
Pressing Utility&Service&User Cal Status displays the summary
results of the previous User Cal, and the status of probe
calibrations for probes that can be calibrated. Note that passive
To Channel 4
6000 Series Oscilloscope User's Guide
To Channel 3
Longer cable
to TRIG OUT

Advertisement

Table of Contents
loading

Table of Contents