Trigger System - Agilent Technologies 54610B User's And Service Manual

Table of Contents

Advertisement

Trigger System

Sources:
Channels 1, 2, line, and external
Internal trigger
1
Sensitivity
:
dc to 25 MHz
100 MHz to 500 MHz
Coupling: ac, dc, LF reject, HF reject, and noise reject
LF reject attenuates signals below 50 kHz, and
HF reject attenuates signals above 50 kHz
Modes: Auto, Autolevel, Normal, Single, and TV
TV triggering: Available on channels 1 and 2
TV line and field: 0.5 division of composite sync for stable display
Holdoff: Adjustable from 200 ns to ≈13 s
External trigger
Range: ±18 V
1
Sensitivity
:
dc to 100 MHz
100 MHz to 500 MHz
Trigger View: External trigger input is displayed along with
channel 1 and channel 2
Trigger View Bandwidth: ≥ 300 MHz
Coupling: ac, dc, LF reject, HF reject, and noise reject
Input resistance: 1 MΩ or 50Ω
Input capacitance: ≈12 pf
Maximum input voltage: 250 V [(dc + peak ac(<10 kHz)]
1
Tested, see "To verify trigger sensitivity," on page 3-17.
0.5 div or 5.0 mV
1 div or 10 mV
<75 mV
<150 mV
Performance Characteristics
Trigger System
4-5