Agilent Technologies MSO6012A User Manual page 160

6000 series
Hide thumbs Also See for MSO6012A:
Table of Contents

Advertisement

N O T E
6000 Series Oscilloscope User's Guide
4 Press the Cyc Ct Bas softkey, and use the Entry knob to select
the cycle count base value from 0 through the Cyc Ct Rep value
minus 1, or All.
5 Press the Segment softkey to select the type of segments from:
Static Segments
Dynamic Segments
Symbol Segments
Idle Segments
6 When Static or Dynamic Segments are selected:
a Press the Slot Id softkey, and use the Entry knob to select
the time slot ID value from All or 1 to 2047. The maximum
static slot value and the minimum dynamic slot value are
limited by the downloaded FIBEX file schedule.
b Press the Slot Type softkey to select the slot type:
All Slots — the oscilloscope triggers on slots regardless of
whether they contain a frame.
Empty Slots — the oscilloscope triggers when a slot does
not contain a frame. Usually used to determine if an
assigned slot is empty when it should have contained a
frame.
Since up to 64 cycles can be defined for FlexRay it may be helpful to set the
oscilloscope to Normal trigger mode instead of Auto trigger mode. This
prevents the oscilloscope from Auto triggering while waiting for a
particular slot in a particular cycle.
Triggering on FlexRay Errors
1 In the FlexRay Trigger Menu (see
on page 157), press the Trigger softkey, and use
Trigger Menu"
the Entry knob to select Error.
2 Press the Errors softkey, and use the Entry knob to select the
error type from:
All Errors
Code Error (NRZ) — NRZ decoding errors only.
Triggering the Oscilloscope