6000L Series Specifications And Characteristics; Performance Characteristics - Agilent Technologies MSO6012A User Manual

6000 series
Hide thumbs Also See for MSO6012A:
Table of Contents

Advertisement

9
Characteristics and Specifications

6000L Series Specifications and Characteristics

Performance characteristics

356
Scope input
Channels
Bandwidth (-3 dB)*
Maximum input
1
Full Scale range
Input impedance
Offset range
Connector
BW limit
Noise peak-to-peak
* Denotes warranted specifications, all others are typical. Specifications are valid
after a 30-minute warm-up period and ±10 ° C from firmware calibration
temperature.
1
1 mV/div is a magnification of 2 mV/div. 2 mV/div is a magnification of 4 mV/div
setting. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div
sensitivity setting and 32 mV for 2 mV/div sensitivity setting.
2
Four 50 Ω termination adapters are supplied with DSO6014L.
Ch 1, 2, 3 and 4 simultaneous acquisition
DSO6014L: DC to 100 MHz
DSO6054L: DC to 500 MHz
DSO6104L: DC to 1 GHz
CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400
Vpk
With 10073C/10074C 10:1 probe: CAT I 500 Vpk,
CAT II 400 Vpk
5 Vrms with 50 Ω input
DSO6014L: 1 mV/div to 5 V/div (1 MΩ)
DSO6054L: 2 mV/div to 5 V/div (1 ΜΩ or 50 Ω)
DSO6104L: 2 mV/div to 5 V/div (1 MΩ), 2 mV/div
to 1 V/div (50 Ω)
DSO6014L2: 1 MΩ ± 1% || 11pF
DSO6054L/6104L: 1 MΩ ± 1% || 14pF or 50 MΩ
± 1.5%, selectable
Coupling AC, DC
±5 V on ranges < 10 mV/div
±20 V on ranges 10 mV/div to 200 mV/div
±75 V on ranges ≥ 200 mV/div
BNC
DSO6014L: 20MHz
DSO6054L/6104L: 25 MHz selectable
DSO6014L: 3% full scale or 2 mV, whichever is
greater
DSO6054L: 3% full scale or 3.6 mV, whichever is
greater
DSO6104L: 3% full scale or 4.5 mV, whichever is
greater
6000 Series Oscilloscope User's Guide