6000 Series Oscilloscope User's Guide
bus is measured. This controls the point within the bit's time
where the bit value is captured.
One Bit
Sample Point
8 Set the trigger level.
For analog channels, turn the Trigger Level knob.
For digital channels, press the D15 Thru D0 key and the
Thresholds softkey to access the threshold level setting
softkeys.
Changing the Source settings in the Serial Decode setup also changes
them in the Trigger setup.
9 Return to the previous menu by pressing the up-arrow
softkey.
10 If the decode line does not appear on the display, press the
Decode softkey to turn it on.
11 If the oscilloscope is stopped, press the Run/Stop key to
acquire and decode data.
If the setup does not produce a stable trigger, the CAN signal may be slow
enough that the oscilloscope is AutoTriggering. Press the Mode/Coupling
key, then press the Mode softkey to set the trigger mode from Auto to
Normal.
You can use the Delayed horizontal sweep mode for easier
navigation of the decoded data.