Making Measurements
242
Phase
Phase is the calculated phase shift from source 1 to source
2, expressed in degrees. Negative phase shift values indicate
that the rising edge of source 1 occurred after the rising
edge of source 2.
Phase =
Source 1
Source 2
1 Press the Settings softkey to select the source 1 and
source 2 channels for the phase measurement.
The default Phase settings measure from channel 1 to
channel 2.
The example below shows a phase measurement between the
channel 1 and the math d/dt function on channel 1.
Figure 40
Delay
---------------------------------------- - X 360
Source 1 Period
Period
Delay
Phase Measurement
6000 Series Oscilloscope User's Guide