Specifications & characteristics
The following tables list the performance specifications and operating characteristics for the
Agilent Technologies 54835A/45A/46A oscilloscope. Asterisks (*) denotes warranted
specifications, all others are typical. Specifications are valid after a 30 minute warm-up period,
±
°
and within
5
C from the self-calibration temperature.
Acquisition
Maximum Sample Rate
Maximum Effective Sample Rate
Memory Depth
Memory Depth Modes
Sampling Modes
Filters
Averaging
Real Time
Equivalent Time
2-channel mode: 65,536 points
4-channel mode: 32,768 points
Auto
Manual
Real Time
Equivalent Time
9-bit Bandwidth Limit filter: BW = (Sample Rate)/20
(Sin x)/x Interpolation: On/Off selectable FIR digital filter. Digital signal
processing adds points between acquired data points to enhance
measurement accuracy and waveform display quality.
BW= Sample Rate/4
Selectable from 2 to 4096.
Chapter 1: General Information
Specifications & characteristics
Agilent Models
54845B — 2-channel mode: 8 GSa/s
54845B — 4-channel mode: 4 GSa/s
54846B — 2-channel mode: 8 GSa/s
54846B — 4-channel mode: 4 GSa/s
500 GSa/s
Agilent Agilent 54845B, and Agilent
54846B
Optimized for best combination of
update rate and display quality.
Selectable
2-channel mode: from 16 to 65,536
points
4-channel mode: from 16 to 32,768
points
Successive single shot acquisitions.
Random repetitive sampling (higher
time resolution at fast sweep speeds).
1–7