14.4.4. *IDN?
This query requests the source to identify itself. It returns the data in four fields separated by
commas.
Query Syntax
Returned Parameters
Example
Note:
The 801RP responds to a *IDN? query with model number 1001P. The 1251RP
responds to a *IDN? query with model number 1251P This is normal and
should be interpreted as a 801RP or 1251RP respectively. There are no
functional differences between the P and RP products with respect to remote
control.
To differentiate between a 1001P or a 801RP, query the current limit setting.
For a 801RP, this value will be 6.0. For a 1001P, this value will be 7.4. There is
no way (an no need) to differentiate a 1251P from a 1251RP over the bus.
14.4.5. *RCL
This command restores the AC source to a state that was previously stored in memory with a
*SAV command to the specified location. All states are recalled with the following exceptions:
·
DISPlay:MODE
Command Syntax
Parameters
Example
Related Commands
66
*IDN?
<AARD>
Field
CI
xxxxxx
nnnnnn
Rev. x.x
"CI,1251P,S123,Rev 1.0"
*RCL <NRf>
0 through 7
*RCL 3
*RST
*SAV
WARNING:
Recalling a previously stored state may place
hazardous voltages at the AC source output.
March 2011
Information
Manufacturer
Model number and letter
Serial number or 0
Revision level of firmware
RP Series