Registration; Figure 9-3: Com Port Properties Dialog Box - Ametek 801RP Series User And Programming Manual

Hide thumbs Also See for 801RP Series:
Table of Contents

Advertisement

1. Close the PGUI32 program if it is still open.
2. Use the Windows Explorer and select the PGUI32 directory or the directory into which
you installed the PGUI32 program.
3. Locate the PGUI32.INI file (Type = Configuration Settings).
4. Open the PGUI32.INI file using a text editor such as Notepad.
5. Locate the [Interface] section and find the entry called "CmdDelay". If no such entry is
present add it yourself.
6. The default value for the CmdDelay is set to 0. Change this value to 10.
7. Your PGUI32.INI file should look like this now:
[Interface]
Bus=0
CommPort=2
GPIBDriver=0
ACSourceAddress=1
CmdDelay=10
[PowerSource]
Model_number=0
[WindowState]
Top=5272
Left=6112
Height=4815
Width=6975
8. Save the file using the File, Save menu.
9. Close Notepad and start the PGUI32.
10. Try to connect again. You may have to
cycle the power on the AC source if it
still shows the Err -100 display.
Higher values than 10 may be set for the
CmdDelay parameter if this fix does not
resolve your problem.