Calibration

Routine calibration should be performed every 12 months. Non-routine calibration is only
required if a related assembly is replaced or if the periodic calibration is unsuccessful.
The calibration involves removing the top cover from the power source. Remove the line
power from the power source before removing the top cover. Some of the adjustments are
on the Oscillator Assembly A4. Other adjustment are on the DC-AC Board Assembly. Refer
to Figure 6-3 for the location of the adjustments.
6.1.

Calibration Equipment

Digital Multimeter:
100 milliohm Current Shunt:
Load Bank:
6.2.

Routine Calibration

Setup:
Connect the test equipment to the power source as shown in Figure 6-1. Refer to Figure 6-3
for adjustment locations.
The DMM can be connected to the front panel connectors for convenience but will have to be
changed when switching voltage ranges. To avoid this, use the common output terminal at
the rear of the unit.
For best results use the output connectors (Front or Rear) that are typically used.
24
Fluke 8840 or HP 34401A
Isotek Model RUG-Z-R100-0.1 or equivalent
Various power load resistors will be needed

Table 2: Load and current

Model
Range
Current
max
801RP
135V
6.0A
801RP
270V
3.0A
1251RP
135V
9.2A
1251RP
270V
4.6A
March 2011
Load
22.5
90.0
14.6
58.6
RP Series