Table 11: Bit Configuration Of Standard Event Status Enable Register - Ametek 801RP Series User And Programming Manual

Hide thumbs Also See for 801RP Series:
Table of Contents

Advertisement

14.4.1. *CLS
This command clears the following registers (see chapter 16 for descriptions of all status
registers):
·
Standard Event Status
·
Status Byte
·
Error Queue
Command Syntax
Parameters
14.4.2. *ESE
This command programs the Standard Event Status Enable register bits. The programming
determines which events of the Standard Event Status Event register (see *ESR?) are
allowed to set the ESB (Event Summary Bit) of the Status Byte register. A "1" in the bit
position enables the corresponding event. All of the enabled events of the Standard Event
Status Enable Register are logically ORed to cause the Event Summary Bit (ESB) of the
Status Byte Register to be set. See Section 16 for descriptions of the Standard Event Status
registers.
The query reads the Standard Event Status Enable register.

Table 11: Bit configuration of standard event status enable register

Bit Position
Bit Name
Bit Weight
CME
EXE
PON
Command Syntax
Parameters
Power-On Value
Example
Query Syntax
Returned Parameters
Related Commands
64
*CLS
None
7
6
5
PON
not
CME
used
128
32
Command error
Execution error
Power-on
*ESE <NRf>
0 - 255
128
*ESE 129
*ESE?
<NR1>(Register value)
*ESR? *STB?
March 2011
4
3
EXE
DDE
QYE
16
8
DDE
Device-dependent error
QYE
Query error
2
1
0
not
not
used
used
4
RP Series

Advertisement

Table of Contents
loading

This manual is also suitable for:

1251rp series

Table of Contents