7
Specifications
Frequency Specifications
Table 7-8 Frequency specifications
Range
99.999 Hz
999.99 Hz
9.9999 kHz
99.999 kHz
999.99 kHz
Notes for frequency specifications:
1 The input signal is lower than the product of 20000000V×Hz (product of voltage & frequency); overload protection: 1000 V.
2 For non-square wave signals, an additional 5 counts need to be added.
3 The multimeter will automatically select the most appropriate range when making frequency measurements.
Duty Cycle and Pulse Width Specifications
Table 7-9 Duty cycle and pulse width specifications
Function
Duty cycle
Notes for duty cycle specifications:
1 The accuracy for duty cycle and pulse width is based on a 5 V square wave input into the DC 5 V range.
2 For AC coupling, the duty cycle range can be measured for signal frequency > 20 Hz.
188
Resolution
0.001 Hz
0.01 Hz
0.0001 kHz
0.001 kHz
0.01 kHz
Mode
DC Coupling
0.01% to 99.99%
AC Coupling
Accuracy
[2]
0.02% + 3
0.02% + 3
< 600 kHz
Range
5% to 95%
Minimum Input Frequency
1 Hz
Resolution
Accuracy (at full scale)
-
0.3% per kHz + 0.3%
-
0.3% per kHz + 0.3%
U1253B User's and Service Guide
[1]