Figure 2.3 Maximum Load Wire Length For 1 V Line Drop - Xantrex XFR 7.5-140 Operating Manual

Xfr 1200 watt series programmable dc power supply
Hide thumbs Also See for XFR 7.5-140:
Table of Contents

Advertisement

Installation
Load Connection
Load Wiring Length for Operation with Sense Lines For applications using
remote sensing, you must limit the voltage drop across each load line. Figure 2.3
shows some maximum allowable wire lengths for a given load current and wire size.
We recommend that you use the larger load wiring to ensure a smaller voltage drop
(1 V typical maximum), although units will compensate for up to 5 V drop in each
line. See also Local and Remote Sensing.
.
100
90
80
70
60
50
40
30
20
10

Figure 2.3 Maximum Load Wire Length for 1 V Line Drop

Noise and Impedance Effects To minimize noise pickup or radiation, use
shielded pair wiring of shortest possible length for load wires. Connect the shield to
the chassis via a rear panel mounting screw. Where shielding is impossible or
impractical, simply twisting the wires together will offer some noise immunity.
When using local sense connections, use the largest practical wire size to minimize
the effects of load line impedance on the regulation of the supply.
46
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
WIRE GAUGE (AWG)
16
12
8
4
14
10
6
LOAD CURRENT (AMPS)
Operating Manual for XFR 1.2kW Series Power Supply
1
2/0
2
1/0