Calibration/Adjustment*

Always calibrate/adjust the balance/scale after setting it up in a new location.
Available Features
Calibration/adjustment can only be performed when
– there is no load on the balance/scale,
– the balance/scale is tared,
– the internal signal is stable.
If these conditions are not met, an error message is displayed.
The weight required for calibration/adjustment is displayed (see "Accessories" for
calibration weights; for model GC503 see "Equipment Supplied").
lite
On Basic
balances, calibration/adjustment can be performed with any of three weight
units: g, kg**, or lb (menu code 1. 4. x)
Factory setting: Grams (menu code 1. 4. 1)
Example
Calibrate the balance/scale
Menu code settings: Factory-set codes
Step
1. Turn on the balance/scale
2. Tare the balance/scale
3. Begin calibration
Calibration weight is displayed
without weight unit (here: 1000 g)
4. Place the indicated calibration
weight on the balance/scale
After calibration, the calibration
weight is displayed with wt. unit
5. Remove the calibration weight
* = "Calibration" technically means to determine the difference between the balance/scale readout
and the actual weight on the pan to determine the accuracy.
Adjustment means to bring a balance/scale into the state of accuracy required for its use.
Therefore, "calibration," as used in this manual, actually means "adjustment."
**= not on models with a readability of 0.1 mg
Key (or instruction) Display/Printout
e
t
t >2 sec.
+
+
0.0 g
1000.0
1000.0
1000.0 g
0.0 g
9