Example
Determine an unknown percentage; store the weight on the balance/scale as a reference
percentage
Settings (changes in the factory settings required for this example):
Menu: Application program: Weighing in percent (menu code 2. 1. 5)
Menu: Reference percentage 100 % (menu code 3. 3. 5)
Step
11. Turn on the balance/scale
12. Tare the balance/scale
13. Display the reference
percentage
14. Place the reference weight
for 100% on the balance/scale
(here: 222.5 g)
15. Start application; if the print
format is set to include data
ID codes,
the following is printed:
16. Place an unknown weight
on the balance/scale
(here: 322.5 g)
17. Display weight
18. Display percentage
19. Unload the balance/scale
10. Delete the reference percentage
11. Repeat the procedure starting
from step 6, if desired.
Key (or instruction) Display/Printout
e
t
v > 2 sec.
+
v
+ 100.00 %
Wxx%
+ 144.94 %
v
+
v
+ 144.94 %
c
0.0 g
ref 100
222.5 g
+
222.500 g
322.5 g
0.00 %
13