12 Technical data

Pos: 28.2 /==== Wechsel ein- auf zweispaltig ==== @ 0\mod_1130421847171_3101.doc @ 3828
Change from one to two columns
Pos: 28.3 /Überschriften/1.1/2-spaltig/D - F/Eingang @ 0\mod_1139395051468_3101.doc @ 3221

12.1 Input

Pos: 28.4 /Technische Daten / Datenblatt/Temperatur/TTR/Technische Daten/Eingang @ 15\mod_1194864822515_3101.doc @ 139745

12.1.1 Resistance

RTD resistance thermometer
Pt100 in acc. with DIN IEC 60751, JIS, MIL, Ni in acc. with
DIN 43760, Cu
(for additional information, see the section "Measuring accuracy")
Resistance measurement
0 ... 500 Ω
0 ... 5000 Ω
Sensor connections
2-, 3-, 4-wire circuit
Connecting cables
2-, 3-, 4-wire max. sensor line resistance (R
in acc. with NE 89 (March 2003);
(3-wire balanced, 2-wire circuit compensation up to 100 Ω sensor
total line resistance)
Measurement current
< 300 µA
Sensor short-circuit
< 5 Ω (for RTD)
Sensor wire break (temperature resistance measurement 2-, 3-,
4-wire)
Measuring range 0 ... 500 Ω
Measuring range 0 ... 5 kΩ
Corrosion detection in accordance with NAMUR NE 89
3-wire resistance reading > 50 Ω
4-wire resistance reading > 50 Ω

12.1.2 Thermocouples/Voltages

Types
B, E, J, K, L, N, R, S, T, U, C, D
Voltages
-125 ... 125 mV
-125 ... 1100 mV
Connecting cables
Max. sensor line resistance (R
Sensor wire break monitoring in accordance with Namur NE 89
pulsed with 1 µA outside the measurement interval
Thermocouple measurement
Voltage measurement
Input resistance
> 10 MΩ
Internal reference junction
Pt1000, DIN IEC 60751 Cl. B
(no jumpers necessary)
Sensor matching
via single point (offset adjustment)
via two point adjustment
Sensor fault signaling
RTD sensor:
Linear resistance measurement:
Thermocouple:
Linear voltage measurement:
Pos: 28.5 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831
OI/TTR200-EN
) for each wire 50 Ω
W
> 0.6 ... 10 kΩ
> 5.3 ... 10 kΩ
) for each line 1.5 kΩ, total 3 kΩ
W
5.3 ... 10 kΩ
5.3 ... 10 kΩ
Short circuit and wire break
Wire break
Wire break
Wire break
Pos: 28.6 /Technische Daten / Datenblatt/Temperatur/TTH/Technische Daten/Ausgang @ 0\mod_1139219931451_3101.doc @ 3765

12.2 Output

Transmission characteristics
temperature linear
resistance linear
voltage linear
Output signal
Configurable 4 ... 20 mA (standard)
Configurable 20 ... 4 mA
(NE43 dynamic range: 3.8 ... 20.5 mA)
Simulation mode
3.5 ... 23.6 mA
Induced current consumption
< 3.5 mA
Maximum output current
23.6 mA
Configurable error current signal
override
22 mA (20.0 ... 23.6 mA)
underdrive
3.6 mA (3.5 ... 4.0 mA)