10-2. Output example

To print the GLP report, set the function to "info 1" and set the AD-8121B printer to
MODE 3.
To output the GLP report to a personal computer, set the function to "info 2".
Data format for "calibration report"
Perform calibration using a weight. Then the balance will output a calibration report.
See "7-1. Calibration using a weight" about the calibration.
AD-8121 format "info 1"
MODEL
S/N
1234567
ID
DATE
09/03/21
02:53:21 PM
CALIBRATED(EXT.)
CAL.WEIGHT
+300.00 g
SIGNATURE
- - - - - - - -
Data format for "calibration test report"
Perform calibration test using a weight. Then the balance will output a calibration test
report. See "10-3. Calibration test using a weight " about the calibration test.
AD-8121 format "info 1"
MODEL
S/N
1234567
ID
DATE
09/03/21
03:15:40 PM
CALIBRATED(EXT.)
ACTUAL
0.00 g
+299.99 g
TARGET
+300.00 g
SIGNATURE
- - - - - - - -
CR :Carriage return, ASCII 0Dh
LF :Line feed, ASCII 04Ah
A & D
Manufacturer
EJ-300
Serial number
ABCDEF
Calibration weight
A & D
Manufacturer
EJ-300
Serial number
ABCDEF
Target weight
:Space, ASCII 20h
Model
ID number
Date
Time
Calibration
type
Signature
Model
ID number
Date
Time
Calibration
test type
Zero value
value
Target weight
Signature
31
General format "info 2"
~~~~~~~~~~~A~&~D<CRLF>
MODEL~~~~~EJ-300<CRLF>
S/N~~~~~~1234567<CRLF>
ID~~~~~~~~ABCDEF<CRLF>
DATE<CRLF>
<CRLF>
TIME<CRLF>
<CRLF>
CALIBRATED(EXT.)<CRLF>
CAL.WEIGHT<CRLF>
~~~~~~+300.00~~g<CRLF>
SIGNATURE<CRLF>
<CRLF>
<CRLF>
-~-~-~-~-~-~-~-~<CRLF>
<CRLF>
<CRLF>
General format "info 2"
~~~~~~~~~~~A~&~D<CRLF>
MODEL~~~~~EJ-300<CRLF>
S/N~~~~~~1234567<CRLF>
ID~~~~~~~~ABCDEF<CRLF>
DATE<CRLF>
<CRLF>
TIME<CRLF>
<CRLF>
CAL.TEST(EXT.)<CRLF>
ACTUAL<CRLF>
~~~~~~~~~0.00~~g<CRLF>
~~~~~~+229.99~~g<CRLF>
TARGET<CRLF>
~~~~~~+300.00~~g<CRLF>
SIGNATURE<CRLF>
<CRLF>
<CRLF>
-~-~-~-~-~-~-~-~<CRLF>
<CRLF>
<CRLF>

Hide quick links: