The results of the comparison are indicated by HI, OK or LO on the display.
The comparison is as follows:
LO < Lower limit value ≤ OK ≤ Upper limit value < HI
Operating conditions (see the function setting "Cp"):
d = the smallest display division
The upper limit and lower limit numerical values are common to each of the weighing,
counting and percent mode. The example for EK-610j is as follows.
Upper limit value "001010": "10.10g" "1010pcs" "101.0%"
Lower limit value "000990":

6-1. Setting example

This example will be "Compares plus data except those
near zero".
Selecting a comparison mode
1. Press and hold the SAMPLE key to display
func
(If the comparison mode is already set, press the
SAMPLE
lower limit values".)
2. Press the ENTER key, then the balance displays
poff X .
3. Press the SAMPLE key several times to display
Cp
X .
4. Press the TARE key several times to display
Cp
3 .
5. Press the ENTER key to store the settings.
Cp Hi
Visit www.scaleman.com to order A&D Weighing balances at discount prices

6. COMPARATOR

No comparison (comparator function disabled).
Compares all data.
Compares all stable data.
Compares plus data except those near zero (plus data greater than +4d).
Compares stable plus data except those near zero (stable plus data
greater than +4d).
Compares all data except those near zero (all data greater than +4d or
less than -4d).
Compares stable data except those near zero (stable data greater than
+4d or less than -4d).
e.g.: 4d=four display divisions
.
key to go to "Entering the upper and
appears after end
"9.90g"
"990pcs"
.
10
"99.0%"
Press and hold
func
poff 0
Press
several times
Cp
0
Press
several times
Cp
3
To store