AND EJ-120 Instruction Manual page 35

Ej series compact balance
Table of Contents

Advertisement

2. CC 0
is displayed.
3. If it is necessary to change the target weight value that is used for calibration test,
press the SAMPLE key and change the value using the following keys.
SAMPLE key
RE-ZERO key
PRINT key
4. At step 2, press the PRINT key. The zero point
is weighed and the weight value is displayed for a
few seconds.
5. Place the target weight with the same value as
displayed on the pan. Press the PRINT key to
weigh it. The weight value is displayed for a few
seconds.
6. end
appears.
7. glp
is displayed and calibration test report is
output.
8. end
appears again. Remove the weight and
press the UNITS key to return to the weighing
mode.
To select the digit blinking to be changed.
To set the value of the digit selected.
To store the value and return to step 2.
The weighed data
is shown with "g".
The weighed data
is shown with "g".
33
CC 0
With nothing
on the pan
CC 0
0. 0
2000. 0
Target
weight
2000. 0
2000. 0
end
glp
GLP output
end
Returns to
the weighing mode