AND EJ-120 Instruction Manual page 17

Ej series compact balance
Table of Contents

Advertisement

5. At step 3, press the
zero-point value. Do not touch the pan during
weighing.
Note
If SPAN calibration is not to be performed, press
the CAL switch or UNITS key to return to the
weighing mode.
6. Place a calibration weight with the same value as
displayed on the pan. Press the
weigh it. Do not touch the pan during weighing.
7. end
appears.
Remove the weight from the pan.
When the GLP output (function setting "info 1" or
"info 2") is selected,
calibration report is output and
again.
Press the CAL switch or
the weighing mode.
Note
The value set in step 4 is stored in memory even after the power is
switched OFF.
If the balance is to be moved to other places, set the gravity
acceleration value of the area where the calibration using a weight is
to be done, and calibrate the balance according to the procedure
above. See the next section to set the value.
PRINT
key to weigh the
PRINT
glp
is displayed. The
end
UNITS
key to return to
15
Cal 0
5000. 0
key to
5000. 0
0end000
appears
glp
0end000
Returns to
the weighing mode
With nothing
on the pan
To store
Calibration
weight
To store
GLP output