50 -
100 -
CAL errors
Cal e
-Cal e
Battery error
lb0
Hb0
AC adapter error
Hb1
lb1
Stability error
error1
If you cannot cancel an error or other errors have occurred, request service from
the store where you purchased the balance or from your local A&D dealer.
Note: Pressing the
samples may reduce the counting accuracy.
Starting from the 100 samples,
be displayed when the sample weight is light.
In this case, press the
adding any samples.
When "aCai 0" (ACAI disabled) or "Umin 2" is
selected, this notice is not shown.
Warning to indicate that calibration has been canceled
because the calibration weight is too heavy.
Warning to indicate that calibration has been canceled
because the calibration weight is too light.
Check the weighing pan and the calibration weight.
To return to the weighing mode, press the UNITS key.
Warning to show that the batteries are exhausted.
Replace them with new one or use the AC adapter.
Warning to show that the battery voltage is too high.
Check the battery.
Warning to show that the output voltage of an AC adapter
is too high. Check if the AC adapter is the correct type.
Warning to show that the output voltage of an AC adapter
is too low. Check if the AC adapter is the correct type.
Warning to indicate that the weight value is not stable
and the balance cannot display it. Prevent vibration and
drafts. Press the UNITS key to return to the weighing
mode.
35
PRINT
key without adding
100 -
PRINT
may
key without