Mechanical

Dimensions
Weight
Mounting
Portability

Electrical

Power Requirements

Battery

Sealed lead acid, with external charger.
The Spot LXi battery is 90 to 100% charged after 6 hours of charging. The rechargeable
batteries in the Braun ThermoScan PRO 4000 thermometer requires an additional 1 hour
to charge. The battery automatically charges when Spot LXi is powered through the AC
power transformer. An operator can use the device while the battery is charging; however,
the battery charges faster when the instrument is not in operation.
The external printer has a Ni-Mh battery that includes a separate external printer charger
(713703 or BC1300M). Charge the external printer battery 6 hours before initial use.

Environmental

Operating Temperature
Storage/Transport Temperature
Relative Humidity
Operating Altitude

Patents

The Spot LXi is covered under the following patents: 6,827 ,488; 6,971,790; and other
patents pending.
WARNING This device is not suitable for use in the presence of a flammable
anesthetic mixture with air or oxygen or nitrous oxide. An explosion may result.
May be hand-carried when held by the rear handle
Patient-rated transformer is connected to AC mains:
100-240V, 50-60Hz, 0.4A or internal power: 6Vdc, 6Ah
Device with SureTemp Plus: -13 to 131F (-25 to 55C)
Device with Braun ThermoScan PRO 4000: -4 to 122F (-20 to 50C)
Specifications
Height: 10.63 " (27 cm)
Length/Braun: 8" (20.32 cm)
Length/SureTemp Plus: 7.5" (19 cm)
Depth: 5.25" (13.34 cm)
7.5 lbs (3.4 kg)
Self-supporting on rubber feet
Custom mobile stand
Custom wall mount
50° to 104° F (10° to 40° C)
15 to 95% (non-condensing)
-557 to 16,000 ft. (-170 to 4877 m)
59