Calibration/Adjustment
You can calibrate/adjust the scale by pressing the = key (4). To do so, menu code 1 5 1 must be
selected. Calibration weight: 5,000 g; accuracy: class F1 + 0.075 g.
After connection to AC power
and before each calibration/
adjustment, allow the scale to
warm up for approx. 30 min.
Formulation Mode (Calculation by a Factor)
This mode enables you to weigh in amounts that are smaller or larger than that of your basic formula
for a specific paint color (e.g., 25 l of a 1-l formula), without having to do any manual recalculation.
You can select from among the following factors (amounts) by pressing the v key (7):
0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0, 10, 15, 20, 25.
As you pour in the components of your formula, the weight is displayed in "g."
Note:
The flashing arrow n on the display means that the weight shown is not valid for use in legal
metrology (not legal for trade).
Weighing with One Decimal Place
Place an empty paint can on the load plate (1).
Press = (4). The display shows "0.0 g."
Pour in the first component, and read off the weight
as soon as the stability symbol appears; in this case, "g."
Pour in additional components until the desired
weight of your formula is reached.
Remove the filled paint can from the load plate.
G Never use a hammer to close the lid of a
paint can while it is still on the load plate, as this
will damage the weighing system.
Hold down the = key (4) for
2 sec. When 5000 is
displayed, release the key.
Center the calibration weight
on the load plate (1).
Calibration/adjustment is
performed automatically. After
calibration/adjustment, remove
the calibration weight.