Adjusting The Back Angle - Invacare XtraTM 1026793 Operator And Maintenance Manual

Invacare wheelchair owner's manual
Table of Contents

Advertisement

Adjusting the Back Angle

NOTE: For this procedure, refer to FIGURE 5.2.
NOTE: This procedure applies to adjustable angle backs only.
NOTE: The back adjusts to four (4) positions in five 5° degree increments from 0° to 15°.
1. Remove the two (2) back angle adjustment screws and locknuts.
2. Reposition the back to the desired angle.
3. Ensure the upper back angle mounting hole is aligned with one (1) of the four (4)
upper angle adjustment holes.
4. Install one (1) of the back angle adjustment screws through the upper back angle
mounting hole and the desired upper angle adjustment hole.
5. Install the remaining back angle adjustment screw through the lower back angle
mounting hole and the desired lower angle adjustment hole.
6. Secure each back angle adjustment screw with a locknut. Torque locknut to 80-100 in/
lbs.
Back Cane
Back Angle
Adjustment
Screw
Locknut
Lower Back
Angle Mounting
Hole
Part No 1026793
Upper Back Angle
Mounting Hole
Locknut
Back Angle
Adjustment
Screw
FIGURE 5.2 Adjusting the Back Angle
35
SECTION 5—BACK
Detail "A" - back
angle mounting holes
Upper
10°
15°
Lower
10°
15°
Xtra™

Advertisement