User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope

How to Implement the Auxiliary System Function Setting

Press the UNTILITY button and the menu is displayed in the screen as Fig.5-40.
The description of the Auxiliary Function Menu is shown as the following table.
Function Menu
System Status
Recall Factory
Do Self Cal
Language
Do Self Cal (Self-Calibration)
The self-calibration procedure can improve the accuracy of the oscilloscope under the
ambient temperature to the greatest extent. If the change of the ambient temperature is
up to or exceeds 5 , the self
highest level of accuracy.
Before executing the self-calibration procedure, disconnect the probe or wire and the
input connector. Then,press the F3 menu selection button and choose "Do Self Cal".
After confirming it is all set, push down F3 button and choose "Do Self Cal",
entering the self-calibration procedure of the instrument.
SYS STAT (System State)
Press the F1 menu selection button and choose "SYS STAT" item. The menu pops up
in the screen as Fig.5-41.
Fig.5-40 Function Menu
Setting
Description
Display the system function menu.
Call out the factory settings.
Carry out the self-calibration procedure.
Chinese
Choose the display language of the operating
English
system.
-calibration procedure should be executed to obtain the
55