5. Care and Maintenance

General Maintenance

Important
If the result of any of these checks is
not satisfactory, discontinue using the
equipment and contact an authorized
technician.
06/2009
To make sure that the system functions correctly, you must have it serviced
annually by an authorized technician. In addition, every three months inspect
the equipment and make sure of the following:
Generator
The certification label is legible.
There are no oil leaks.
Mechanical support
The wall framework is securely attached to the wall.
All the labels are legible.
The scissor arm is stable in all positions.
Control unit and electrical installation
The symbols are legible.
The control unit cable and the power supply cable are in good
condition.
The ground is correctly installed.
The radiology control key returns to its initial position after use.
Functioning
The audible signal is audible and the x-ray emission light is visible
when you make an exposure (manual mode, 70kV, 0.1 sec.).
The message "E 01", which means Operator Error, is displayed when
you make an exposure (manual mode, 70kV, 1.0 sec.) and release the
control button before the exposure time has elapsed.
Timer self-test
Turn on the system to activate the self-test.
The test starts with a simultaneous test of the display and alarm
lights.
The unit proceeds to the systems test. At the end of this test,
indicated by a short beep, the firmware version and the total number
of exposures (divided by 10) made by the machine since installation
is displayed.
If the test is not successful, an Error code is displayed on the display.
SM731_K2200_01_en