After Using The Defibrillator - HP Codemaster 100 User Manual

Portable defibrillator/monitor
Table of Contents

Advertisement

After Using the Defibrillator

If you want to print an Event Summary now, press Review. (See "Printing the
Advisory Event Summary Record" on page 4-16.) After you use the defibrillator,
perform the following steps to prepare the defibrillator for its next use.
o
Clean all paddles, controls, and cables. Visually check the patient cables,
paddles, paddles cables, and pads adapter cables for wear, insulation nicks, and
other damage. Refer to Chapter 13, "Maintenance" for additional information.
o
Turn on the HP CodeMaster 100. If the instrument displays LOW BATTERY,
replace the battery pack with a charged battery.
o
Perform the test described in "Delivered Energy and Shock Button Functional
Test" on page 13-3.
o
Turn the Energy Select control to Off and return the HP CodeMaster 100 to its
storage location.
o
Check that sufficient recorder paper and electrolyte medium or multifunction
defib electrodes are available for the next use of the defibrillator.
o
Make sure that a battery is in the battery support system being charged. See
"Charging the Battery" on page 13-14.