Defibrillation During Pacing
If the patient must be defibrillated during pacing, perform the following steps.
1
Set the desired energy level with the Energy Select control.
2
Press Charge. The defibrillator will automatically turn off the pacer and start
charging. The pacer status messages will be cleared and replaced with the
defibrillator delivered energy display.
3
Call out "Clear!" to alert personnel to stand away from the patient.
4
Wait for the Charge Done Indicators.
5
Simultaneously press and briefly hold both Shock (
the cable connector. The shock will be delivered through the multifunction defib
electrodes.
After the shock, the pacer remains off. Press the Start/Stop button to resume pacing
if it is required.
Using the Pacer
) buttons, located on
8-7