Cleaning Exterior Surfaces On The Ac And Dc Power Modules; Maintaining The Battery - HP Codemaster 100 User Manual

Portable defibrillator/monitor
Table of Contents

Advertisement

CodeMaster 100

Cleaning Exterior Surfaces on the AC and DC Power Modules

Maintaining the Battery

Use only the HP M2480B battery support system to charge
HP M2476B/M2477B battery packs. You should also use only the HP M2480B
battery support system to discharge, recondition and perform a capacity test
on these batteries. It is recommended that you charge batteries at room
temperature (10-30°C).
Be sure to alternate batteries between the HP CodeMaster 100 and the
battery support system. Install a freshly charged battery pack in the HP
CodeMaster 100 whenever the unit displays the warning message LOW
BATTERY. Return the depleted battery pack to the battery support system to
be charged.
Perform the capacity test on each battery every 90 days or when the battery
support system indicates a capacity test should be performed. If capacity test
fails, remove the battery pack from service.
Use batteries regularly. If stored longer than 30 days, perform a capacity test
on the battery before using it.
13-11