Heart Rate Alarms

Heart Rate Alarms
5
The HP CodeMaster 100 provides three configurable pairs of upper and lower
heart rate alarm limits. Each pair of heart rate alarm limits can be defined as
described in Table 2-1 on page 2-11. While monitoring, you can select and
enable any pair of pre-defined limits using HR Alarm.
When heart rate alarms are inactive, the monitor will display
.
To select a pair of heart rate alarm limits, press HR Alarm until the pair of
limits you wish to use are displayed. If you do not press the key again, the
displayed heart rate alarm limits become active and the limits are replaced by
the bell symbol.
If the heart rate alarm limits are violated, the heart rate alarm limits replace
the bell symbol, and the violated limit is highlighted. Pressing HR Alarm at
this point will turn off the heart rate alarms.
If the heart rate alarms are active and you wish to review the limits, press HR
Alarm. The currently active pair of heart rate alarm limits are displayed
momentarily.
If the heart rate alarms are active and you wish to select another pair of
limits, press HR Alarm until the pair of limits you wish to use are displayed.
Pressing HR Alarm repeatedly cycles through the three pairs of heart rate
alarm limits and the heart rate alarms inactive choice.
Heart rate alarms are automatically turned off when you press Charge.
NOTE
5-9