Printing the Advisory Event Summary Record
Figure 4-5
Advisory Event Summary Record
4-5
To print the advisory event summary record, press Review.
To stop printing the advisory event summary, press Review or Record.
To review the advisory event summary after the instrument has been turned off or is
turned to manual mode, turn the Energy Select control to Advisory On and press
Review.
The header record lists when power on occurred, when the last event occurred, and
how many shocks were delivered. The record also includes areas for you to write in
the patient's name, the operator's name, and any comments about the event.
The directory lists all events that occurred during the resuscitation attempt and the
time of their occurrence.
4-17