After Using The Defibrillator; Defibrillating With Alternate Paddle Sets - HP CodeMaster XL+ M1722A User Manual

Defibrillator/monitor
Table of Contents

Advertisement

Defibrillating
Defibrillating a Patient

After Using the Defibrillator

If you want to print an Event Summary now, press
Event Summary Record on page 3-8. After you use the defibrillator, perform the
following steps to prepare the defibrillator for its next use.
1
Turn the Energy Select control to Off (Standby).
2
Return the instrument to its storage location, and plug the power cord into an AC
power outlet. Verify that the
3
Clean all paddles, controls, and cables. Refer to Chapter 8, Maintaining the
Defibrillator for cleaning instructions.
4
Check that sufficient recorder paper and electrolyte paste or defibrillator pads
are available for the next use of the defibrillator.

Defibrillating with Alternate Paddle Sets

The CodeMaster XL+ will defibrillate with several different pads/paddles sets.
• Adult/Pediatric Anterior/Anterior External Paddles
• External Adhesive Pads
• Internal Paddles
Performing Pediatric Defibrillation
The CodeMaster XL+ paddle set comes with pediatric paddles. To use the pediatric
paddle set, depress the release latch at the front of the standard external paddle set
while pulling forward on the adult paddle surface. This action will remove the adult
paddle contact surface and expose the smaller pediatric contact surface.
Refer to the Defibrillating a Patient section for defibrillation procedures.
2-4
Review
BATT
and [
AC
. See Printing the
lights are on.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Codemaster xl+ m1722b

Table of Contents