12-Lead (Option C90)
Basic ECG Controls:
ID, 12-Lead, and Transmit, ON, Copy, and Filter Keys, Alphanumeric QWERTY
keyboard
Signal Processing/Recording:
Frequency response: Selectable from these ranges;
0.05-150, 0.15-150,
0.05-100, 0.15-100,
0.05-40,
0.15-40,
When 0.15-150Hz is selected, the 12-Lead recorder tracing meets the Frequency
and Impulse Response specification of ANSI/AAMI EC11-1991 Diagnostic
Electrocardiographic Devices by methods A, D and E.
Filters:
Noise Filter: On/Off capability. Power line filter always enabled. Configurable
baseline wander, noise, and artifact filters.
Storage:
Capacity: 30 ECGs
Type: 2 MB of non-removable flash memory
Transmission of 12-Lead ECGs to HP ECG Manager (PC-Based Receiving
Station Software):
-Internal cellular/landline modem with RJ11C connector (US, Canada, UK only)
-RS232 Port for external modem.
Modem Command Interfaces:
Data: Hayes standard AT command set
Fax: EIA/TIA - 578 Service Class I
Recommended Modem Protocols:
-Modulation Protocol: V.34
-Error Correction Protocol: V.42
-Compression Protocol: V.42 bis
-FAX Modulation Protocol: V.17, V.29, V.27ter, V.21 channel 2
0.5-150Hz
0.5-100Hz
0.5-40Hz
Defibrillator/Monitor