Performing Cardioversion
There are three ways to monitor ECG for synchronized cardioversion: use an
external ECG monitor; connect ECG electrodes to the HP CodeMaster 100; or use
multifunction defib electrodes.
External ECG monitor When the patient is already connected to bedside
monitoring equipment, there is an external monitoring cable which plugs into the
ECG Output jack of the bedside monitor and connects to the HP CodeMaster 100
ECG Input for monitoring. To use an external monitor, perform the following steps:
1
Select Lead I or Lead II on the HP CodeMaster 100.
2
Plug the cable into the external monitor ECG Output jack and plug the input end
of the cable into the ECG input plug on the HP CodeMaster 100.
Using ECG electrodes with the HP CodeMaster 100 Use a patient cable connected
to the patient and select Lead I, Lead II, or Lead III, choosing the best lead that
displays a large QRS complex.
Using multifunction defib electrodes Use the electrode adapter cable (M2472A)
and defib electrodes, selecting PADS as the ECG source.