CodeMaster 100
Keep hands clear of the paddle electrode edges. Use your thumbs to
WARNING
depress the Shock buttons on the paddle handles.
10 Grasp the paddle handles, and without removing the paddles from their
11 Press Sync to place the instrument in Sync mode.
12 Verify that the messages SYNC and USE LEADS appear on the display.
13 Press Lead Select once to select Lead I, and verify that the message USE
14 Turn the Energy Select control to Off.
The defibrillator is ready for use if it passes the above checklist.
A more extensive test of defibrillator/monitor functionality can be performed
using the diagnostic service mode described in the HP M2475B CodeMaster
100 Portable Defibrillator/Monitor Service Manual.
12-12
holders, press both Shock buttons simultaneously. A brief automatic
recorder run prints the delivered energy test report.
LEADS no longer appears. The message LEADS OFF should now appear,
indicating that one or more leads are not connected.