Delivered Energy and Shock Button Functional Test

To check the HP CodeMaster 100 with the paddles, perform the following
steps:
The HP CodeMaster 100 is designed to protect against delivering a shock
NOTE
when only one switch on the paddle set is pushed.
1 Turn the Energy Select control to the 100 joules position.
2 Verify that the adult paddle electrodes are installed.
3 Push the paddles completely into their holders (Apex paddle in right
Keep hands clear of the paddle electrode edges. Use your thumbs to
WARNING
depress the Shock (
4 With the paddles in their holders, grasp the paddle handles and press the
5 Release the Apex paddle Shock button, then press the Sternum paddle
6 Press Sync to place the defibrillator in sync mode.
7 Press and hold both Shock buttons. Verify that the defibrillator does not
8 Press Sync again to remove the defibrillator from sync mode.
9 With the paddles in their holders, press and briefly hold both Shock but-
10 The recorder will print a test report. Review the test report for proper
pocket, Sternum in left) and press either Charge. Wait for the Charge
Done indicators.
) buttons on the paddle handles.
Apex paddle Shock (
discharge.
Shock button. Verify that the defibrillator does not discharge.
discharge.
tons at once. Verify that the defibrillator discharges.
operation.
) button. Verify that the defibrillator does not
CodeMaster 100
13-3