Setting Up Filters

From the Configuration Menu, select Setup ECG Filters. The 12-Lead option has
been factory-configured with the filter settings which remove the most noise from
the ECG. In addition to the default settings, 12-Lead offers a choice of several filter
configurations. These filter settings are detailed below and in Table 2-5.
12-Lead Report Filters -
Wander Filter:
Noise Filter:
Front Panel Filter Key-
Muscle Artifact Filter:
40 Hz Noise Filter:
0.5 Hz Wander Filter:
F1
F2
• The 0.5 Hz Baseline Wander filter suppresses the greatest amount of baseline
wander.
• The 0.15 Hz Baseline Wander filter provides some baseline wander suppression
without distorting the ECG's ST segment.
• The 0.05 Hz Baseline Wander filter delivers the highest fidelity signal, but pro-
vides the least baseline wander suppression.
• The 40 Hz Noise filter offers maximum noise suppression, but reduces the fidelity
of the signal.
• The 100 Hz Noise filter provides some noise suppression while offering an accu-
rate signal representation.
• The 150 Hz Noise filter delivers the highest fidelity signal, but provides the least
high-frequency noise suppression.
Setup ECG Filters
Change Value
F3
Configuring the 12-Lead Option
1.5 Hz
40 Hz
Yes
Yes
Yes
Exit
F4
F5
2-19