Advisory Mode Default Settings

Advisory Mode Default Settings
4
The HP CodeMaster 100 contains default settings for use in Advisory mode. Some
of the settings can be altered using the Setup Menu.
The following settings can be changed:
• Default energy select sequence. You can choose to change the sequence from the
factory default settings for a sequence of 200, 200, and 360 Joules, to a sequence
of 200, 300, and 360 Joules. The energy level for the third shock and all subse-
quent shocks will be 360 Joules.
• Post Shock Data: On/Off.
When operating in Advisory mode, the following settings cannot be changed:
• Charge done tone: always sounds.
• CRT alert tones: always sounds.
• Alert tone volume: always at the maximum volume.
• Recorder delay: no delay.
• Recorder Bandwidth: always Monitor bandwidth.
• Automatic Recordings: always off.
• Power on Lead: always PADS/PADDLES.
• Notch Filter: always on.