HP CodeMaster 100 and 12-Lead Option Interactions

HP CodeMaster 100 and 12-Lead Option Interactions
6
This section describes the interactions that occur between the HP
CodeMaster 100 and the 12-Lead option.
• If the HP CodeMaster 100 is reset or a shock is delivered while the 12-Lead
option is attempting to acquire or store an ECG, 12-Lead operation is sus-
pended and the ECG will be lost (with no recovery).
The HP CodeMaster 100 resets when the following occurs:
-delivering a shock
-changing modes (manual to shock advisory)
-changing the battery
To resume transmitting after the unit is reset, select the ECG from the Send
ECGs Menu and press Send.
• The 12-Lead option does not operate while the HP CodeMaster 100 is set to
Shock Advisory mode.
• Low Battery information from the HP CodeMaster 100 displays on the
12-Lead screen.
• When the language option is changed in SETUP MENU 1 on the HP Code-
Master 100, the 12-Lead user interface will display the new language upon
the next power on .
• Since the HP CodeMaster 100 and 12-Lead option share the 50mm recorder,
any action which would normally cause the HP CodeMaster 100 to print will
interrupt any report printing on the 12-Lead. Press Try Again to resume
12-Lead printing.
6-34