Specifications

14

Defibrillator/Monitor

14
Waveform: Damped sinusoidal (Lown).
Output Energy (Delivered): 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, and
360 joules.
Charge Control: Push-button on apex paddle and on front panel.
Charge Time: Less than 5 seconds to 360 joules with a fully charged battery
installed. Time required to charge fully discharged energy storage device to
maximum energy (360 joules)
Armed Indicators: Charge done tone, charge done lamp on apex paddle, and
available energy indicated on display.
Paddle Contact Indicator: 3-color LED bar graph array on STERNUM paddle
indicates quality of defibrillator paddle contact before discharge.
Paddles: Standard paddles are water-resistant, anterior/anterior, adult and pediatric
with PCI Contact indicator. External paddles are water resistant. Adult electrodes
(83 cm sq.) slide off to expose pediatric electrodes (21 cm sq.). Paddle cord is 10 ft
(3 m). Full range of internal paddles available.
Synchronizer: SYNC message appears on monitor and is annotated periodically on
recorder while in synchronous mode. An audible beep sounds with each detected R-
wave, while a marker on the monitor and sync designator on the recorder strip
indicate the discharge point.
Environmental Operating Conditions: 0 to 55°C, 15 to 95% relative humidity
(noncondensing) for 24 hours at 40°C, 15,000 ft altitude. (Readability of the 12-
Lead display may be degraded outside the range of 0 to 40°C.)
Environmental Storage Conditions: -20 to 70°C, 90% relative humidity
(noncondensing) for 24 hours at 65°C, 15,000 ft altitude. (See also Battery Storage
Specification.)