Performing Cardioversion
Using Multifunction Defib Electrodes for Performing
Synchronized Cardioversion
This section provides details for performing synchronized cardioversion with
multifunction defib electrodes. Multifunction defib electrodes can be selected as the
ECG source for cardioversion because they provide a reliable contact for
monitoring.

Applying the Multifunction Defib Electrodes

1
Attach the electrode adapter cable (M2472A) to the defibrillator.
2
Attach the multifunction defib electrodes to the patient as instructed on the elec-
trode package.
3
Connect the electrode cable connector to the electrode adapter cable. The elec-
trodes are correctly connected when the locking ring is twisted, locking the ears
of the connector to the adapter cable.
4
Turn the Energy Select control to Monitor On.
If the PADS OFF monitor message is displayed, check all patient connections.
5
Select pads as the ECG source by pressing Lead Select.
6
Press Sync once to place the HP CodeMaster 100 in Sync mode. The message
SYNC will appear on the display.