5
The ECG output provides an analog 1V/mV ECG signal for connection to a
Hewlett-Packard external monitor. Hewlett-Packard compatible external
monitoring divider cables are as follows.
Table 5-8
External Monitoring Cables
1000:1 Voltage Divider Cable Connector Type
The ECG output is also compatible with the interface to the HP Viridia
NOTE
Information Center and other HP central stations.
Do not use the ECG output to synchronize another defibrillator. (The ECG-In
NOTE
to ECG-Out delay is 35 milliseconds.)
The connection of external equipment may increase leakage currents.
WARNING
Always request that local safety personnel verify that multiple connected
equipment comply with local regulatory standards before putting such
equipment into service.