Display settings
Adjust the slider until there is almost no visible difference
3.
between the background of the image above the slider and the
bitmap inside this image.
When done, click button Set point 1 of 20.
4.
Repeat this procedure until the button is grayed. This indicates the
5.
last point is set.
Click OK to finish.
6.
To reset to the latest calibration:
Follow step 2 and 3 of the calibration with Barco LCD sensor (see
1.
page 77).
In the Calibration dialog (see step no. 4. in the description of the
2.
calibration with sensor), check the option Use the latest
calibration.
Click the OK button.
3.
The calibration and the luminance will be reset to the latest
calibrated position.
Note:
80
Figure 34: DICOM Optimizer dialog
Nio MKII