  • Page 3 Welcome! Nio is Barco's flexible, industry-standard display solution for diagnostic imaging. Presenting an exceptional combination of innovation and performance, Barco's next-generation Nio display system improves your diagnostic reading routine for a multitude of medical imaging applications. Following symbols may be used throughout this guide:...
  • Page 4 (This page intentionally left blank.)
  • Page 5: Table Of Contents

    Table of contents What's in the box.............7 Parts, controls and connectors ........8 Front view ....................8 Rear view ....................9 Display installation ............12 Unlocking the height mechanism ............12 Adjusting the display position ............. 13 Connecting the signal cables............... 16 Routing the cables................
  • Page 6 Concepts..................35 Display function selection ............36 ALC & DICOM Options..............37 Switching DPMS on/off................ 40 Cleaning your display ............41 Front glass ..................... 41 Cabinet....................42 Important information ..........43 Safety information................43 Environmental information..............46 Regulatory compliance information............ 49 Explanation of symbols ................ 51 Legal disclaimer ..................
  • Page 7: What's In The Box

    • If you ordered a Barco display controller, it’s also in the box together with its accessories. A dedicated user guide is available on the system Keep your original packaging. It is designed for this display and...
  • Page 8: Parts, Controls And Connectors

    Parts, controls and connectors Front view Figure 1: Front side Power LED USB downstream port Control wheel...
  • Page 9: Rear View

    Rear view E-3620, MDNG-5121 EAR VIEW OF Figure 2: Rear side Connector compartment Tilt & swivel foot cover cover Tilt & swivel foot Slot for security cable DVI (digital) video input USB upstream port USB downstream port DC power input Tilt &...
  • Page 10 MDNC-3121, MDNG-2121 MDNG-6121 EAR VIEW OF Figure 3: Rear side Connector compartment Tilt & swivel foot cover cover Tilt & swivel foot DC power input +5 Vdc Slot for security cable DVI (digital) video input USB upstream port USB downstream port Tilt &...
  • Page 11 MDNC-2121 EAR VIEW OF Figure 4: Rear side Connector compartment Tilt & swivel foot cover cover Tilt & swivel foot Slot for security cable DisplayPort video input DVI (digital) video input USB upstream port USB downstream port DC power input Tilt &...
  • Page 12: Display Installation

    Display and connecting all ARCO necessary cables, make sure to have a suitable display controller physically installed in your computer. If you are using a Barco display controller, please consult the dedicated user guide available on the system CD. Unlocking the height mechanism In the factory, the height-positioning system in the display foot is blocked with a red clip to prevent damage during transportation.
  • Page 13: Adjusting The Display Position

    Adjusting the display position You can change the orientation of the panel at any time, but it is more convenient to select landscape or portrait orientation before connecting the cables.. Figure 6: Portrait orientation Figure 7: Landscape orientation...
  • Page 14 O CHANGE THE PANEL ORIENTATION Stand at the front side of the panel and take the panel at both sides. Very important: Til the panel before changing the orientation. Should you change the panel orientation without tilting it first, you might irreversibly damage the tilt & swivel mechanism. Figure 8: Tilt the panel before rotating...
  • Page 15 To change from portrait to landscape, turn the panel counterclockwise. Figure 9: To rotate the panel from portrait to landscape To change from landscape to portrait, turn clockwise.. If, after installing the display or the system, you change the panel orientation while an image is on the screen, the result depends on the graphic board and the resolution of the image.
  • Page 16: Connecting The Signal Cables

    Connecting the signal cables O CONNECT THE SIGNAL CABLES TO THE DISPLAY To get access to the connectors, remove the connector compartment cover by pulling down the 2 clips at the top of the cover. The location of the connectors depends on the display type: Figure 10: Inputs E-3620, MDNG-5121 Figure 11: Inputs MDNC-3121, MDNG-2121 and MDNG-6121 Figure 12: Inputs MDNC-2121...
  • Page 17: Routing The Cables

    Routing the cables Bind the cables in the connector compartment together with the • cable tie inside the connector compartment. Put the connector compartment cover back on the display. Pay • attention that the signal cables are positioned under the bulge in the cover.
  • Page 18: Vesa-Mount Installation

    Slide the cover upward while moving the lower side of the cover towards the foot. Press the cover to the foot so that it makes a clicking sound. VESA-mount installation The panel, standard attached to the tilt & swivel foot, is compatible with the VESA 100 mm standard.
  • Page 19 Remove the small screw (A) fixing the small plastic cover on top of the foot. Next, remove the small cover itself. Figure 14: Display with tilt & swivel foot cover removed Unscrew the 2 screws fixing the round plastic cover (B). Lift up the round plastic cover.
  • Page 20: Daily Operation

    DPMS is enabled on your display, but it also needs to be activated on your workstation. To do this, go to “Power Options Properties” in the “Control Panel”. Barco recommends setting DPMS activation after 20 minutes of non-usage. SE A SCREEN SAVER TO AVOID IMAGE RETENTION Prolonged operation of an LCD with the same content on the same screen area may result in a form of image retention.
  • Page 21 NHANCE USER COMFORT Every Barco multi-head display system is color matched with the highest specification in the market. Barco recommends keeping color-matched displays together. Furthermore, it is important to use all displays of a multi-head configuration at the same rate to preserve color matching throughout the economic lifetime of the system.
  • Page 22: Stand-By Switching

    Stand-by switching The connected power supply also provides a switch that can be used to turn the power completely off. To use the display, please make sure to switch on this power supply. This can be done by pushing the on/off switch on the power supply into the "|"...
  • Page 23: Locking And Unlocking User Controls

    MDNC-2121, MDNC-3121, MDNG-2121 MDNG- 6121 DISPLAYS When the display is on and no on-screen display is visible, push the control wheel at the front shortly. An on-screen message appears, asking to press to confirm. Push the control wheel once more to confirm and switch to stand- The LED turns orange.
  • Page 24 E-3620 MDNG-5121: O DISABLE USER CONTROLS ON Rotate the control wheel to display the on-screen display. The Main Menu appears. Rotate the control wheel to select Settings. Press the control wheel to enter the Settings menu. SETTINGS DPMS Power LED User Controls EXIT Figure 18: Settings menu (example)
  • Page 25: Bringing Up The Osd Menu

    Steps 4 to 6 must be performed in maximum 3 seconds. In MDNC-2121, MDNC-3121, MDNG-2121 and MDNG-6121 displays, the user controls are automatically enabled after this procedure. In other Nio displays, you must enable the user controls again in the Settings menu.
  • Page 26: Navigating Through The Osd Menus

    Navigating through the OSD menus Rotate the control wheel to move up and down through the menus. To go into a submenu, rotate the control wheel to highlight the submenu. Next, press the control wheel to go into the submenu. When you press on Exit, the OSD exits the submenu and jumps to the previous (higher-level) menu.
  • Page 27: Changing Values In The Osd Menus

    Changing values in the OSD menus Navigate through the menus and highlight the parameter you wish to adjust. Press the control wheel to go into edit mode. Rotate the wheel to change the value. Press the wheel to enter the new value and return to the menu.
  • Page 28: Advanced Operation

    Advanced operation On-screen display (OSD) Input Selection menu* Auto Detects the incoming video signal automatically and displays it on the screen. Always displays the DVI signal on the sreen. When no DVI signal available, there will be no image on the screen. Always displays the DisplayPort signal on the screen.
  • Page 29: Luminance And Color Menu

    Luminance and color menu In grayscale Nio displays this menu is called “Luminance”. Measured Indicates the actual luminance measured luminance** by the internal sensor. This is a read-only value. It is expressed in a percentage of the calibrated value (100% is calibrated).
  • Page 30: Display Function Menu

    When you change the luminance target, the display will adjust its backlight to reach the target. This can be seen in the Measured luminance line. When the luminance target cannot be reached, e.g., due to aging of the backlight, the Measured luminance line changes to Minimum value reached or Maximum value reached.
  • Page 31 ALC & DICOM O PTIONS SUBMENU * Correction value Shows the ambient light correction value that is taken into account in the calculation of the display function, measured in lux. DICOM offset The DICOM DF is recalculated at the moment the DICOM Offset is changed, taking the new DICOM Offset into account.
  • Page 32 EADING ROOM DEFINITION SUBMENU * Reading room Indicates the reading room type you are editing. You can select another room to edit by rotating the control wheel. Max. ambient light This indicates the maximum ambient light that corresponds to the selected reading room.
  • Page 33: Settings Menu

    Settings menu DPMS Allows to switch the display power management system on/off. See note below. Power LED Allows to switch the power LED’s on state on/off. The LED's orange DPMS state is not influenced by this setting. So, when the display goes into power-saving mode, the LED will turn orange, even if it was switched off by this setting...
  • Page 34: Information Menu

    Backlight Runtime* Indicates the total time the display has been operating, excluding the time in stand-by In some NIO displays these functions are called “Display Lifetime” and “Backlight Lifetime”. Their function, however, is the same as described in the table above.
  • Page 35: Changing Display Functions

    Changing Display Functions Concepts Display function selection is possible only in MDNC-3121, MDNG-2121 and MDNG-6121 displays. The Display Function menu lets you select a display function (DF). • When you select the DICOM DF you can define the DICOM DF to be •...
  • Page 36: Display Function Selection

    Display function selection (DF): O SELECT A DISPLAY FUNCTION When there is no OSD on the screen, rotate the control wheel to display the OSD. The main menu appears. Rotate the control wheel to select the Display Function menu. Press the control wheel shortly to enter the menu. Rotate the control wheel to select Display Function.
  • Page 37: Alc & Dicom Options

    User This is a user-definable DF. Use MediCal QAWeb to define the DF. ALC & DICOM Options DICOM offset NTRODUCTION The DICOM DF will be recalculated only when another DICOM Offset is selected. The DICOM DF will be recalculated based on the new offset setting.
  • Page 38: Reading Room Selection

    OU CAN SELECT THE FOLLOWING OFFSETS Dark room The DICOM DF will be recalculated taking an ambient light value of 0 lux into account. This means that the measured ambient light does not influence the DICOM DF. Preset The DICOM DF will be recalculated taking a preset ambient light value into account.
  • Page 39: Reading Room Definition

    CT/MR/NM Corresponds to light conditions in diagnostic reading rooms for computed tomography, magnetic resonance or nuclear medicine scans. Office Corresponds to light conditions in office rooms. Clinical Corresponds to light conditions in diagnostic reading rooms for clinical viewing. Emergency Corresponds to light conditions in emergency rooms.
  • Page 40: Switching Dpms On/Off

    To modify the maximum ambient light for this reading room, rotate the control wheel to select Max. Ambient Light. Press the control wheel shortly to go into edit mode. Rotate the control wheel to change the value. Press the control wheel shortly to save the changes.
  • Page 41: Cleaning Your Display

    Cleaning your display Front glass Clean the front glass using a sponge, cleaning cloth or soft tissue, lightly moistened with one of the following tested products: Flux Pril • • Windex Glass Plus Ajax glass cleaner • • Bohle glass cleaner Sidolin glass cleaner •...
  • Page 42: Cabinet

    Do not use following products: Alcohol/solvents at higher concentration > 5% • Strong alkalis lye, strong solvents • Acid • Detergents with fluoride • Detergents with ammonia • Detergents with abrasives • Steel wool • Sponge with abrasives • Steel blades •...
  • Page 43: Important Information

    Important information Safety information ENERAL ECOMMENDATIONS Read the safety and operating instructions before operating the equipment. Retain safety and operating instructions for future reference. Adhere to all warnings on the equipment and in the operating instructions manual. Follow all instructions for operation and use. LECTRICAL SHOCK YPE OF PROTECTION ELECTRICAL...
  • Page 44 12 VDC OWER CONNECTION QUIPMENT WITH EXTERNAL POWER SUPPLY Power requirements: The equipment must be powered using the • delivered medical approved 12 VDC power supply. The medical approved DC power supply must be powered by the • AC mains voltage (protective earth terminal). 24 VDC OWER CONNECTION QUIPMENT WITH EXTERNAL...
  • Page 45 ATER AND MOISTURE Never expose the equipment to rain or moisture. Never use the equipment near water - e.g. near a bathtub, washbasin, swimming pool, kitchen sink, laundry tub or in a wet basement. ENTILATION Do not cover or block the ventilation openings in the cover of the set. When installing the equipment in a cupboard or another closed location, heed the necessary space between the set and the sides of the cupboard.
  • Page 46: Environmental Information

    ERCURY NOTICE This Barco product consists of materials that may contain mercury, which must be recycled or disposed of in accordance with local, state, or country laws: Within this system, the backlight lamps in the monitor display •...
  • Page 47 Electronic Information Products" (Also called RoHS of Chinese Mainland), the table below lists the names and contents of toxic and/or hazardous substances that Barco's product may contain. The RoHS of Chinese Mainland is included in the MCV standard of the Ministry of Information Industry of China, in the section "Limit Requirements of toxic substances...
  • Page 48 Pollution Control Labeling Standard" of Chinese Mainland, marked with the Environmental Friendly Use Period (EFUP) logo. The number inside the EFUP logo that Barco uses (please refer to the photo) is based on the "Standard of Electronic Information Products Environmental Friendly Use...
  • Page 49: Regulatory Compliance Information

    Regulatory compliance information : E-3620, MDNG-2121, MDNC-3121, NDICATIONS FOR USE MDNC-2121 The products are intended to be used in displaying and viewing digital images, for review and analysis by trained medical practitioners. These devices must not be used in primary image diagnosis in mammography.
  • Page 50 Reorient or relocate the receiving antenna. • Increase the separation between the equipment and receiver. • Connect the equipment into an outlet on a circuit different from • that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help. •...
  • Page 51: Explanation Of Symbols

    Explanation of symbols On the display or power supply, you may find the following symbols (nonrestrictive list): Indicates compliance to the essential requirements of the Directive 93/42/EEC 0120 Indicates compliance with Part 15 of the FCC rules (Class A or Class B) Indicates the display is approved according to the UL regulations Indicates the display is approved according...
  • Page 52: Legal Disclaimer

    Barco software products are the property of Barco. They are distributed under copyright by Barco N.V. or Barco, Inc., for use only under the specific terms of a software license agreement between Barco N.V. or Barco Inc.
  • Page 53: Technical Specifications

    Technical specifications Product acronym MDNG-2121 Screen technology TFT AM LCD Dual Domain IPS Active screen size 540 mm (21.3") (diagonal) Active screen size (H x V) 432 x 324 mm (17.0 x 12.8") Aspect ratio (H:V) Resolution 2MP (1600 x 1200) Pixel pitch 0.2700 mm Color imaging...
  • Page 54 Response time (Tr + Tf) 23 ms Scanning frequency (H; V) 15-129 kHz; 24-100 Hz Housing color Gray Video input signals DVI-D Dual Link USB ports 1 upstream (endpoint), 2 downstream USB standard Power requirements 100-240V (nominal) Power consumption (nominal) Power save mode Power management DVI-DMPM...
  • Page 55 Swivel -45° / +45° Pivot 90° Mounting standard VESA (100 mm) Screen protection Protective, non-reflective PMMA cover Recommended modalities CT, MR, US, DR, CR, NM, Film Certifications BSMI, CCC - GB4943-2001 - GB9254-1998 - GB9254- 2008 - GB17625.1-2003, DEMKO EN60601-1:1990 + A1:1993 + A2:1995, IEC 60601-1:1998 + A1:91 + A2:95, UL 60601-1 1st edition, CAN/CSA-CS22.2 No.601.1-M90, FDA510K, RoHs, MDD, WEEE,CE 0120,...
  • Page 56 Product acronym E-3620 (MA) Screen technology TFT AM LCD Dual Domain IPS Active screen size 528 mm (20.8") (diagonal) Active screen size (H x V) 424 x 318 mm (16.7 x 12.5") Aspect ratio (H:V) Resolution 3MP (2048 x 1536) Pixel pitch 0.207 mm Color imaging...
  • Page 57 Housing color Gray Video input signals DVI-D Dual Link USB ports 1 upstream, 2 downstream USB standard Power requirements 100-240V (nominal) Power consumption (nominal) Power save mode Power management DVI-DMPM Dot clock 280 MHz OSD languages English Dimensions with stand (W Portrait: 385 x 525~585 x 250 mm Landscape: 485 x 475~535 x 250 mm x H x D)
  • Page 58 Mounting standard VESA (100 mm) Screen protection Protective, non-reflective PMMA cover Recommended modalities CT, MR, US, DR, CR, NM, Film Certifications CCC - GB4943-2001 - GB9254-1998 (Class A) - GB9254-2008 (Class A) - GB17625.1-2003, DEMKO EN60601-1:1990 + A1:1993 + A2:1995 + A11: 1993 + A12: 1993 + A13:1996, IEC 60601-1:1988 + A1:1991 + A2:1995 2nd edition, UL 60601-1 1st edition 2006-04- 26, CAN/CSA-C22.2 No.601.1-M90 2005, FDA510K,...
  • Page 59 Product acronym MDNG-5121 Screen technology TFT AM LCD Dual Domain IPS Active screen size 541 mm (21.3") (diagonal) Active screen size (H x V) 422 x 338 mm (16.5 x 13.3") Aspect ratio (H:V) Resolution 5MP (2560 x 2048) Pixel pitch 0.165 mm Color imaging Gray imaging...
  • Page 60 Housing color Gray Video input signals DVI-D Dual Link USB ports 1 upstream (endpoint), 2 downstream USB standard Power requirements 100-250V (nominal) Power consumption (nominal) Power save mode Power management DVI-DMPM Dot clock 330 MHz OSD languages English Dimensions with stand (W Portrait: 408 x 489~549 x 250 mm Landscape: 492 x 531~591 x 250 mm x H x D)
  • Page 61 Mounting standard VESA (100 mm) Screen protection Protective, non-reflective PMMA cover Recommended modalities CT, MR, US, DR, CR, NM, Film Certifications BSMI, CCC - GB4943-2001 - GB9254-1998 (Class A) - GB9254-2008 (Class A) - GB17625.1-2003, DEMKO EN60601-1:1990 + A1:1993 + A2:1995, IEC 60601- 1:1998 + A1:1991 + A2:1995 2nd edition, IEC 60601-1 (ed.2) am1 &...
  • Page 62 3MP HB OLOR Product acronym MDNC-3121 Screen technology TFT AM Color LCD IPS Active screen size 540 mm (21.3") (diagonal) Active screen size (H x V) 433 x 325 mm (17.0 x 12.8") Aspect ratio (H:V) Resolution 3MP (2048 x 1536) Pixel pitch 0.2115 mm Color imaging...
  • Page 63 Video input signals DVI-D Dual Link USB ports 1 upstream (endpoint), 2 downstream USB standard Power requirements 100-250V (nominal) Power consumption (nominal) Power save mode Power management DVI-DMPM Dot clock 165 MHz OSD languages English, French, German, Spanish, Italian Dimensions with stand (W Portrait: 385 x 525~585 x 250 mm Landscape: 485 x 475~535 x 250 mm x H x D)
  • Page 64 Screen protection Protective, non-reflective PMMA cover Recommended modalities CT, MR, US, DR, CR, NM, Film Certifications BSMI, CCC - GB4943-2001 - GB9254-2008 - GB17625.1-2003, DEMKO EN60601-1:1990 + A1:1993 + A2:1995 + A13:1996, IEC 60601-1:1988 + A1:1991 + A2:1995 2nd edition, UL 60601-1 1st edition 2006-04- 26, CAN/CSA-C22.2 No.601.1-M90 2005, FDA510K, RoHs, MDD, WEEE, CE 0120, VCCI, FCC Class B, KETI, IEC 60601-1:2005, EN 60601-1-2:2006, ANSI/AAMI ES...
  • Page 65 Product acronym MDNG-6121 Screen technology TFT AM LCD Dual Domain IPS Active screen size 541 mm (21.3") (diagonal) Active screen size (H x V) 432,6 x 324,5 mm (17.0 x 12.8") Aspect ratio (H:V) Resolution 5MP (2800 x 2096) Pixel pitch 0.1540 mm Color imaging Gray imaging...
  • Page 66 Housing color Gray Video input signals DVI-D Dual Link USB ports 1 upstream (endpoint), 2 downstream USB standard Power requirements 100-240V (nominal) Power consumption (nominal) Power save mode Power management DVI-DMPM Dot clock 280 MHz OSD languages English, French, German, Spanish, Italian Dimensions with stand (W Portrait: 385 x 525~585 x 250 mm Landscape: 485 x 475~535 x 250 mm...
  • Page 67 Mounting standard VESA (100 mm) Screen protection Protective, non-reflective PMMA cover Recommended modalities CT, MR, US, DR, CR, NM, Film Certifications BSMI, CCC - GB4943-2001 - GB9254-1998 - GB9254- 2008 - GB17625.1-2003, DEMKO EN60601-1:1990 + A1:1993 + A2:1995 + A13:1996, IEC 60601-1:1988 + A1:1991 + A2:1995 2nd edition, UL 60601-1 1st edition 2006-04-26, CAN/CSA-C22.2 No.601.1-M90 2005, FDA510K, RoHs, MDD, WEEE, CE 0120, VCCI, FCC...
  • Page 68 2MP HB OLOR Product acronym MDNC-2121 Screen technology TFT AM Color LCD IPS Active screen size 541 mm (21.3") (diagonal) Active screen size (H x V) 432 x 324 mm (17.0 x 12.8") Aspect ratio (H:V) Resolution 2MP (1600 x 1200) Pixel pitch 0.2700 mm (0.01063”) Color imaging...
  • Page 69 Video input signals DVI-D Single Link (1x), DisplayPort (1x) USB ports 1 upstream (endpoint), 2 downstream USB standard Power source Input for 24 VDC power supply unit: 90-264 VAC Input for display: 24 VDC For use only power: Sinpro HPU100-108 Power consumption (nominal) Power save mode...
  • Page 70 Screen protection Protective, non-reflective PMMA cover Recommended modalities CT, MR, US, DR, CR, NM, Film Certifications UL60601-1, CB60601-1, CSA C22.2 No 601.1, EN60601- 1-2, EN55011 Limit B, ICES-001 B, EN61000-3-2/3, EN61000-4-2/3/4/5/6/8/11, VCCI, CE0120, KETI, CCC, BSMI, Demko, FCCB, FDA 510K, CMDCAS, MDD, WEEE, RoHs, IEC 60601-1:2005, EN 60601-1-2:2006, ANSI/AAMI ES 60601-1:2005, CAN/CSA-C22.2 No.
  • Page 72 B410582-09 November 2012 Barco n.v. President Kennedypark 35 8500 Kortrijk Belgium 0120 www.barco.com...