Display settings
Application Settings dialog is set. See "Calibration tab" in the
description of the Application Settings dialog.
If you have multiple displays connected (multi-head system),
2.
select the display you wish to calibrate by clicking on the
corresponding numbered icon in the Display Settings dialog.
Click the Calibrate button.
3.
A message appears, showing the different calibration options.
4.
Select "Using the BARCO LCD sensor" and click OK.
5.
This option is grayed in case NioWatch does not find a connected
Barco LCD sensor.
Follow the guidelines on the screen to complete the calibration
6.
successfully.
Note:
If the Equalization option is set in the Application Settings and
NioWatch finds at least one other display of the same type that is
calibrated, the "Match with reference" page appears during the
calibration process.
78
Figure 32: Calibration choices
Nio MKII