Fluke 5700A Series II Service Manual page 84

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

5700A/5720A Series II Calibrator
Service Manual
8. Press O to activate the calibrator output. Then turn the output adjustment knob on
9. Se the rear panel CALIBRATION switch to ENABLE. Press E on the front
10. The range calibration is now complete. Set the rear panel CALIBRATION switch to
3-14
PREV
MENU
the calibrator until you achieve a null on the null detector.
panel. The calibrator will now calculate a new range constant multiplier for the 220V
dc range, and will store it in non-volatile memory.
the NORMAL position, disconnect the external standards, and press r to reset the
calibrator to its newly calibrated ranges.
IF057.EPS

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

5720a series ii5720a series ii

Table of Contents