Ac Current Test 22 Ma To 11A Ranges - Fluke 5700A Series II Service Manual

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

5700A/5720A Series II Output
Level
2 mV
20 mV
200 mV
2V
2.3V
20V
200V
300V
600V
1000V
*Adjustment counts of 3 times the listed value is allowed at 1 MHz

3-20. AC Current Test 22 mA to 11A Ranges

This test requires the use of equipment listed in Table 3-9, when using the 5790A Input 1.
Use Tables 3-33 (5720A) or 3-34 (5700A) for a test record.
Table 3-9. Equipment Required for 22 mA to 11A Alternating Current Test Using the 5790A Input 1
Equipment
AC Measurement standard
Current Shunts
Current Shunt Adapter
Cable
Cable (For 10A Setup)
Cable (For 10A Setup)
Adapter (For 10A Setup)
Adapter (For 10A Setup)
Adapter (For 10A Setup)
Table 3-8. 5790A Adjustment Counts
5790A Display
2.0000
20.0000
200.0000
2.000000
2.30000
20.00000
200.0000
300.000
600.000
1000.000
Fluke 5790A
Fluke A40 Series: 20 mA, 200 mA, 2A, and A40A Series: 10A (if verifying a
5725A) , with AC-DC difference corrections.
Fluke 792A-7004 A40 current shunt adapter
Pomona 1368-A-18. Double banana to single banana plugs.
Pomona 5268-C-12. BNC(M) to single banana plugs.
Fluke A45-4003 (PN 212853) UHF(M) to UHF(M) with RG8A/U cable
Pomona Model 1707. UHF(F) to banana adapter.
Kings KC-99-34 UHF(M) to BNC(F)
Pomona Model 1707. UHF(F) to banana adapter.
Calibration and Verification
5790A Adjustment Counts*
Model or Description
Full Verification
1
1
4
4
1
4
4
1
2
2
3
3-31