Fluke 5700A Series II Service Manual page 80

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

5700A/5720A Series II Calibrator
Service Manual
RESISTANCE STANDARD
CURRENT
HI
LO
10. To reject the changes, return to the calibration menu shown in step 2 by pressing
3-10
SENSE
HI
LO
CHASSIS
GROUND
Figure 3-2. 742A-1 and 742A-10k External Calibration Connections
P. Otherwise, press the softkey under "Proceed" to accept and save the changes,
and to open the display shown below, letting you continue with calibration.
PREV
MENU
CALIBRATOR
OUTPUT
SENSE
V A
V
WIDEBAND
HI
HI
LO
LO
HI
AUX
GUARD GROUND
CURRENT
F7-4.EPS
IF3-12.EPS